Analytical HPLC

Analytical HPLC System LAHS-B10 is a unit equipped with LCD display UV-Visible wavelength range of 190 nm to 700 nm and double plunger reciprocating pump system for constant flow pump. It has isocratic kit system for test result accuracy. Consists of separate workstation, sample injector and front mounted self-locking switch. Deuterium lamp is used as light source with valve mounting frame. It is ideally used in the detection of compounds in pharmaceutical industries, research institute, food industries, and petroleum industries, etc.

Read More

Categories: ,

Specifications

High Pressure Constant Flow Pump
Double plunger reciprocating pump 1 set
Flow range 0.001 to 10.00 mL/min
Setting step size 0.001 mL/min
Flow accuracy ≤ ± 0.3 %
Flow stability RSD ≤ 00.1 %
Maximum pressure 42 MPa
Display pressure error ≤ 1.0 %
Pressure fluctuation ≤ 1.0 %
Power consumption 80 W
UV/Visible Detector
Wavelength range 190 to 700 nm
Wavelength repeatability ± 0.1 nm
Band width 8 nm
Response time 0 to 4.9 seconds
Linearity range 1.8 AU (5 %)
Baseline Noise ≤ ± 1 × 10-5AU
Baseline drift ≤ 2.0 × 10-4AU/h
Minimum detectable concentration 4.0 × 10-9 g/mL
Detector cell Optical path: 10 mm


Volume: 10 µL

Light source Deuterium lamp
Power consumption 100 W
Voltage AC 110 V/ 220 V ± 10 %, 50/60 Hz
Packing Dimensions (W × D × H) 600 × 440 × 380 mm


2 × (600 × 410 × 300 mm

Gross weight 54 kg